| 单道手动可调移液器 | |||||||
| 目录号 | 量程(μL) | 增量(μL) | 测量体积(μL) | 允许最大系统误差 | 允许最大随机误差 | ||
| (不准确度) | (不精确度) | ||||||
| % | μL | % | μL | ||||
| SPA200125 | 0.1-2.5 | 0.01 | 2.5 | 2.5 | 0.0625 | 1.6 | 0.04 | 
| 1.25 | 3 | 0.0375 | 3 | 0.0375 | |||
| 0.25 | 12 | 0.03 | 6 | 0.015 | |||
| SPA200510 | 0.5-10 | 0.1 | 10 | 1 | 0.1 | 0.8 | 0.08 | 
| 5 | 2 | 0.1 | 1 | 0.05 | |||
| 1 | 2.5 | 0.025 | 1.5 | 0.015 | |||
| SPA200220 | 2-20 | 0.5 | 20 | 0.9 | 0.18 | 0.4 | 0.08 | 
| 10 | 1.5 | 0.15 | 1 | 0.1 | |||
| 2 | 3 | 0.06 | 2 | 0.04 | |||
| SPA200550 | 5-50 | 1 | 50 | 0.6 | 0.3 | 0.3 | 0.15 | 
| 25 | 0.8 | 0.2 | 0.4 | 0.1 | |||
| 5 | 2 | 0.1 | 2 | 0.1 | |||
| SPA210100 | 10-100 | 0.5 | 100 | 0.8 | 0.8 | 0.15 | 0.15 | 
| 50 | 1 | 0.5 | 0.5 | 0.25 | |||
| 10 | 3 | 0.3 | 1.5 | 0.15 | |||
| SPA220200 | 20-200 | 1 | 200 | 0.6 | 1.2 | 0.15 | 0.3 | 
| 100 | 0.7 | 0.7 | 0.3 | 0.3 | |||
| 20 | 2 | 0.4 | 0.8 | 0.16 | |||
| SPA211000 | 100-1000 | 5 | 1000 | 0.6 | 6 | 0.2 | 2 | 
| 500 | 1 | 5 | 0.4 | 2 | |||
| 100 | 2 | 2 | 0.7 | 0.7 | |||
| 手动固定移液器 | |||||||
| 目录号 | 量程(μL) | 增量(μL) | 测量体积(μL) | 允许最大系统误差 | 允许最大随机误差 | ||
| (不准确度) | (不精确度) | ||||||
| % | μL | % | μL | ||||
| SPA100005 | 5 | - | 5 | 1.3 | 0.065 | 1.2 | 0.06 | 
| SPA100010 | 10 | - | 10 | 0.8 | 0.08 | 0.8 | 0.08 | 
| SPA100050 | 50 | - | 50 | 0.5 | 0.25 | 0.3 | 0.15 | 
| SPA100100 | 100 | - | 100 | 0.5 | 0.5 | 0.3 | 0.3 | 
| SPA100500 | 500 | - | 500 | 0.3 | 1.5 | 0.2 | 1 | 
| SPA101000 | 1000 | - | 1000 | 0.3 | 3 | 0.15 | 1.5 |